August 9, 2016 NEMC Anaheim, California Rock J. Vitale, CEAC; Michael R. Green, Ph.D.; David A. Gratson, CEAC Environmental Standards Inc. W. James Bover, Ph.D. WJB Consulting LLC J. Andrew Tachovsky, P.E. Sabine River Partners Quality Systems for Hydrocarbon Analyses: Similarities and Differences between EPA, ASTM and GPA methods – Considerations for a more Unified Approach Topics in Shale Gas

Upload: david-gratson-ceac

Post on 23-Feb-2017

26 views

Category:

Documents


2 download

TRANSCRIPT

Page 1: Vitale NEMC 2016 Presentation

August 9, 2016 NEMC Anaheim, California

Rock J. Vitale, CEAC; Michael R. Green, Ph.D.; David A. Gratson, CEAC Environmental Standards Inc.

W. James Bover, Ph.D. WJB Consulting LLC

J. Andrew Tachovsky, P.E. Sabine River Partners

Quality Systems for Hydrocarbon Analyses: Similarities and Differences between EPA, ASTM and GPA methods – Considerations for a more Unified Approach

Topics in Shale Gas

Page 2: Vitale NEMC 2016 Presentation

Agenda

CCR Regulations

Current Issues Impacting Oil/Gas Analysis

Elements of Quality System

Overview of EPA, ASTM, GPA Methods

Unification

Page 3: Vitale NEMC 2016 Presentation

Recent Developments in Hydrocarbon Testing

Tier III Vehicle Emission and Fuel Standards - 40 CFR

Part 80, Subpart D

Tier III Revisions: Testing of

reformulated gasoline. As of January 2016

require sampling and subsequent

analysis per §80.47

Change from prescriptive ASTM

test method to §80.47:

Performance-based Analytical Test

Method

Page 4: Vitale NEMC 2016 Presentation

Flash Gas Emission Estimates (e.g. Section 114 request of E&P operators)

Analysis of pressurized

sample (condensate, oil, water) e.g. CARB

Appendix B* Methods include ASTM (e.g. D2597 );

GPA (e.g. 2177, 2286); and U S EPA (e.g. Method TO-

15, 8260) EPA, ASTM, and GPA all

have separate QC Practices

Page 5: Vitale NEMC 2016 Presentation

Crude by Rail - Notice of Proposed Rulemaking

49 CFR Parts 130, 171, 173, and 174

Pipeline and Hazardous Materials Safety Administration (PHMSA), DOT.

Alternatives to D86 for initial boiling point offered.

Page 6: Vitale NEMC 2016 Presentation

Elements of a Quality System

Qualify Employees and Laboratory -

general, not method specific

Train Employees

Qualify method within

laboratory

• Accuracy and precision

On-going demonstration of acceptability for each instrument

and method

• Accuracy and precision

Page 7: Vitale NEMC 2016 Presentation

Quality System Terms

EPA: Blank, LFB/LCS, Duplicates, Matrix Spikes, PE Sample, Surrogates, Internal Standards

GPA: Repeatability, Reproducibility, Reference standard blend, Linearity Plot, Fidelity Plots

ASTM: Repeatability, Reproducibility, SQC, check standard, internal standards, proficiency test samples, verification samples, CRM

Shall/Should/Recommend

Incorporation by Reference

Page 8: Vitale NEMC 2016 Presentation

GPA Quality Practices

Fidelity of Reference Standards (initial and on-going comparison with historic

know standards)

Determination of Linear Range (includes linearity review)

Control Charts (20 separate days, establish limits, outlier tests)

Calibration (single and multi-point, certified standard, verify repeatability, monitor

regularly)

As stated in GPA Methods 2198, 2261

Page 9: Vitale NEMC 2016 Presentation

US EPA Quality Practices

Training and Qualification of employees and

laboratory

Calibration

Batch QC (e.g. blank, LCS, MS)

Sensitivity checks (MDL/RL)

General Requirements

Method (Program) specific QC Proficiency Testing required with some programs & state accreditation RCRA includes performance-based allowances

Page 10: Vitale NEMC 2016 Presentation

US EPA Quality Practices Consider Methods 8260 and

TO-15 (cited by CARB) continued

• Demonstration of Method Capability • TO-15 has sensitivity, precision, and

accuracy limits - method defined. • 8260 has generic starting limits, with

recommendation for in-house derivation. Not based on inter-laboratory studies.

• Initial Calibration (linearity/sensitivity) • Operating Conditions

Page 11: Vitale NEMC 2016 Presentation

US EPA Quality Practices Consider Methods 8260 and

TO-15 (cited by CARB) • Batch Quality Control

• Daily calibration • Blank (negative control) • Internal Standards/Surrogates • Sensitivity (MDL, RL) • Precision (duplicates) & Accuracy (check stds)

assessment • Qualitative Criteria • Method Performance Information (no required

use) • Single and multi-laboratory precision and

accuracy

Page 12: Vitale NEMC 2016 Presentation

The NELAC Institute Environmental Laboratory Sector - Technical Requirements (2009)

• Quantification formulae • Qualitative/selectivity • Constant/reproducible test conditions • Reagents and standards quality • Calibration • Measurement traceability • Batch Quality Control

• Positive & negative control • Precision • Accuracy • Sensitivity (RL, MDL)

• PT Studies for Accreditation

Page 13: Vitale NEMC 2016 Presentation

ASTM Quality Practices – D2

D6792 QA System– Technical Requirements

• Calibration (test method specific) & Verification (CRM)

• Quality Control Practices • Control charts, per D6299 • Accuracy, precision • “Regular” analysis of QC samples (9%, then

per TPI & PT results) • QC based on all technicians and instruments

in lab • QC sample: stable/homogeneous/similar to

samples.

Page 14: Vitale NEMC 2016 Presentation

ASTM Quality Practices – D2

D6792 QA System– Technical Requirements

• Repeatability: single operator, instrument, back to back

• Reproducibility: between laboratories (personnel + instruments)

• Test Performance Index (TPI): compare laboratory precision with published reproducibility • TPI = Test Method reproducibility/site

precision Site precision should be < test method (inter-laboratory) reproducibility

Page 15: Vitale NEMC 2016 Presentation

ASTM Quality Practices System – D2

D6299 Statistical Quality Assurance and Control Charting Techniques to Evaluate Analytical

Measurement System Performance

• Provides practices for quality control and control charting • Reference Material Analysis • Stability and Precision QC Testing • Accuracy Monitoring • Charting Proficiency Testing Data • Periodic, Independent System Validation

(Validation Audit)

Note, there are ASTM standards for general practices included in E260 and E355 (gas chromatography).

Page 16: Vitale NEMC 2016 Presentation

ASTM D6299 and 40 CFR 80.47

• ASTM D6299 was incorporated by reference into 40 CFR 80.47. As such, compliance with both ASTM D6299 and 40 CFR 80.47 may expect to generate the same statistical evaluations:

• I charts

• Outlier evaluation

Page 17: Vitale NEMC 2016 Presentation

ASTM D6299 and 40 CFR 80.47 - continued

• Moving range charts

• Outlier evaluation

Page 18: Vitale NEMC 2016 Presentation

ASTM D6299 and 40 CFR 80.47 - continued

• Normality evaluation

• Q/Q Plots

• Anderson Darling statistics

• Run rule evaluation

• EWMA trend line

calculation and evaluation

Page 19: Vitale NEMC 2016 Presentation

ASTM D6299 and GPA 2198

• GPA 2189 speaks to the same types of control charting practices as D6299, although with less specificity. For example

• 20 points required for statistical assessment • Equivalent methods for calculating standard deviation (site precision) • I charting is explicit in D6299, implied in GPA 2198

• There are other aspects of ASTM D6299 control charting that are not addressed in GPA 2189. For example:

• Normality assessment • Moving range chart preparation • Pre-treatment of results • F-test for similar variances and t-test

for means

Page 20: Vitale NEMC 2016 Presentation

ASTM D6299 and GPA 2198 - Continued

• There are other aspects of ASTM D6299 control charting that are not addressed in GPA 2189. For example:

• Methods for identifying outliers • Calculation of trend is not discussed • Transition to next QC material (e.g.,

Q-procedure) • Allows use of EWMA vice run rules

• In addition, run rule evaluation criteria in D6299 and GPA 2198 are slightly different. D6299 has 4 criteria, whereas GPA 2198 has 6 criteria. Differences include:

• Consecutive points above/below centerline: 9 (D6299) vs. 7 (GPA 2198) • Consecutive points increasing/decreasing: 7 (D6299) vs. 5 (GPA 2198)

Page 21: Vitale NEMC 2016 Presentation

Cross Program Practices EPA GPA ASTM

Calibration Initial and ongoing with method defined

criteria

Fidelity plots for standards, calibration plots for “linearity.”

Test method specific

Negative Control Field, laboratory blanks compared

relative to detection/reporting

limits

Not discussed Some require blanks

Accuracy Control Calibration verification check

standard

Check Standard

Accuracy Control Surrogates, internal standards

Not applicable Internal standards

Accuracy Control Matrix spike Nothing equivalent Nothing equivalent

Accuracy Control PT Control charts Control charts, SQC of reference material,

inter-laboratory cross check

Page 22: Vitale NEMC 2016 Presentation

Cross Program Practices

EPA GPA ASTM

Precision Control Batch duplicates (samples, spikes),

RPD criteria in methods or lab

derived

Daily sample duplicates,

repeatability criteria in methods

SQC of samples, repeatability and

reproducibility criteria in methods

Precision Control Daily calibration standard duplicate,

reproducibility criteria in methods

SQC of reference material,

repeatability and reproducibility

criteria in methods

Page 23: Vitale NEMC 2016 Presentation

Precision Comparison

ASTM & GPA

Repeatability (within laboratory). Difference at 95%

confidence level

Reproducibility (between laboratories). Difference at 95% confidence limit.

ASTM

Precision Ratio (PR) = Reproducibility/Repeatability (integer) • Relative significant difference between

single analysts versus multiple analyst and laboratories.

Test Performance Index (TPI) • Compare individual laboratory with

published reproducibility of standard test method

• F test

Frequency of QC varies with method TPI once established Site Precision

EPA/NELAC

Batch (20 analyses) replicate

comparison (~repeatability as relative percent difference RPD),

Reproducibility not included

Page 24: Vitale NEMC 2016 Presentation

Hydrocarbon Testing Unification

ASTM GPA Tier III (EPA)

Embedded QC Practices -control charting -check standards - positive & negative controls -internal standards -surrogates

Surrogates NA

• Internal Standards

• negative control

• Surrogates NA

Reference to a specific petroleum lab quality system practices

None

None within EPA methods

Specific/strict initial calibration requirements

None

Interlaboratory accuracy/precision checks

None

PT/SRM Program Within some programs

Page 25: Vitale NEMC 2016 Presentation

Thank You

Headquarters 1140 Valley Forge Road | PO Box 810 | Valley Forge, PA 19482 | 610.935.5577

Virginia 1412 Sachem Place, Suite 100 | Charlottesville, VA 22901 | 434.293.4039 Tennessee 8331 East Walker Springs Lane, Suite 402 | Knoxville, TN 37923 | 865.376.7590

New Mexico PO Box 29432 | Santa Fe, NM 87592 | 505.660.8521 Illinois PO Box 335 | Geneva, IL 60134 | 630.262.3979

Web www.envstd.com | E-mail [email protected]

“Setting the Standards for Innovative Environmental Solutions”