    ATOMIC SPECTROMETRY

    Directions

    Universal QTA Burner Mount

    Part Number B300-0350

    For Atomic Absorption Spectrometers

    a rb dc

    Return to Document Menu

  • 7/28/2019 Soporte Universal Fias

    2/11

    Notice

    The information contained in this document is subject to change without notice.

    Release history

    Any comments about the documetation for this product should be addressed to:

    User Assistance

    PerkinElmer, Inc.

    710 Bridgeport Avenue

    Shelton, Connecticut 06484-4794

    U.S.A.

    Or emailed to [email protected]

    Notices

    The information contained in this document is subject to change without notice. Except as specifically set forth in its

    terms and conditions of sale, PerkinElmer makes no warranty of any kind with regaed to this document, including,

    but not limited to, the implied warranties of merchantability and fitness for a particular purpose. PerkinElmer shall

    not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance

    or use of this material.

    Copyright information

    This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be reproduced in any form whatsoever

    or translated into any language without the prior written permission of PerkinElmer, Inc.

    or one of its affiliates.

    Copyright 1999 by affiliates of PerkinElmer, Inc.

    Produced in Singapore

    Trademarks

    PerkinElmer is a trademark of PerkinElmer, Inc.

    AAnalyst trademark of affiliates of PerkinElmer, Inc.

    Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are not to be consid-

    ered unprotected by law.

    Part Number Release Publication Date

    0993-5222 A

    B

    C

    February 1995

    September 1996

    November 1999 Valid for Software Version 6.5

  • 7/28/2019 Soporte Universal Fias

    3/11

    0993-5222 3

    Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Installing the QTA burner mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Remove the burner head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Attach the QTA burner mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Align the QTA burner mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6

    Reinstall the burner head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7

    Installing the heat shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8

    QTAs for FIAS systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9

    Extinguishing the flame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Spectrometers with automatic atomizer changeover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10

    Universal QTA Burner Mount

    Marker for header

    Contents page

  • 7/28/2019 Soporte Universal Fias

    4/114 0993-5222

  • 7/28/2019 Soporte Universal Fias

    5/11

    0993-52225

    Description

    The PerkinElmer Universal QTA Burner Mount is designed to mount the flame-

    heated QTA (QTA=quartz tube atomizer) for MHS-10, MHS-15, and FIAS

    systems on PerkinElmer atomic absorption spectrometers.

    The burner mount installs on the burner head of the AA spectrometer. The QTA

    rests in a hinged bracket. When the bracket is in the down position the QTA is

    located in the spectrometers sample beam and is heated by the flame. When the

    bracket is in the up position the QTA is removed from the sample beam and is not

    heated.

    Note that the QTA is referred to as the quartz cell or the FIAS-cell in some

    publications.

    Figure 1. Universal QTA burner mount (shown with MHS-10/15 QTA)

    Sample inlet

    Hinged QTA bracket(shown in down position)

    Burner head

    Burner spraychamber

    QTA

  • 7/28/2019 Soporte Universal Fias

    6/11

    60993-5222

    Installing the QTA burner mount

    To install the QTA burner mount you must remove the burner head, attach the

    burner mount to the burner head, and then reinstall the burner head.

    Remove the burner head

    1. Where applicable, extinguish the flame, allow the burner head to cool, shut

    down the gases at source, and bleed the gas supply lines (the correct procedure

    is described in the Users Guide for your AA spectrometer).

    2. Remove the burner head as described in the Users Guide for your AA

    spectrometer.

    Attach the QTA burner mount

    Note: A heat shield is provided with the QTA burner mount. You only requirethe heat shield if your spectrometer is a Model 3100, 3110, or 3300 (see page 8).

    1. Swing the QTA bracket on the burner mount into the up position.

    2. Slacken (but do not remove) the two adjusting screws in each securing clamp.

    3. Slide the QTA burner mount onto the burner head and position the two

    securing clamps over the ends of the burner head. The QTA bracket must be

    at the front of the burner head.4. Tighten the adjusting screws, but leave them slack enough so that you can

    move the burner mount.

    Align the QTA burner mount

    1. Swing the QTA bracket into the down position.

    2. Visually sight along the burner slot and move the QTA burner mount with

    respect to the burner head until the two Vs in the QTA bracket line up withthe burner slot.

    3. Tighten the adjusting screws in the securing clamps.

    W4.2

    Warning: High Temperatures Risk of Burns

    The burner head can reach very high temperatures.

    Do not touch the burner head until it has cooled to room temperature.

  • 7/28/2019 Soporte Universal Fias

    7/11

    0993-5222 7

    Figure 2. Attaching the QTA burner mount to the burner head

    Figure 3. Aligning the QTA burner mount on the burner head

    Reinstall the burner head

    1. Reinstall the burner head as described in the Users Guide for your AA

    spectrometer.

    Note: The QTA burner mount can remain permanently installed on the burner

    head.

    If you wish to change over to normal flame analyses, tilt the QTA bracket to the

    up position and park the QTA in the holder at the side of the MHS-10/15 reactor

    assembly. Move the reactor assembly away from the immediate vicinity of the

    atomizer compartment.

    QTA bracketin up position

    Securing clamp(1 of 2)

    Adjusting screw(2 each clamp)

    Burner head

    Burner head

    seen from the

    rear

    Burner slot

    V in QTAbracket

    QTA bracketin down position

  • 7/28/2019 Soporte Universal Fias

    8/11

    8 0993-5222

    Installing the heat shield

    If your spectrometer is a Model 3100, 3110, or 3300, you should attach the heat

    shield to the burner mount after you have installed the burner mount and

    reinstalled the burner head. Refer to Figure 4 while performing the procedure.

    Figure 4. Installing the heat shield (Models 3100, 3110, 3300 only)

    1. Put the QTA bracket in either the down or the up position.

    2. Slide the heat shield onto the QTA bracket so that the slots on the sides of the

    heat shield fit on the shoulder screws at each side of the QTA bracket.

    Shoulder screw

    Heat shield

  • 7/28/2019 Soporte Universal Fias

    9/11

    0993-5222 9

    QTAs for FIAS systems

    If you are using the flame to heat the QTA for a FIAS system, use only the QTA

    with removable PTFE window assemblies. This QTA has a uniform diameter

    along its whole length when the window assemblies are removed. Always remove

    the window assemblies before heating the QTA.

    Do not use a QTA with flared ends and integral exhaust vents.

    FIAS QTA suitable forheating with a flame

    FIAS QTA unsuitable forheating with a flame

  • 7/28/2019 Soporte Universal Fias

    10/11

    10 0993-5222

    Extinguishing the flame

    When you extinguish the flame after the system has been in operation for several

    hours, the QTA is hot and continues to radiate energy until it has cooled down.

    Depending on the model of spectrometer, this energy can be detected by the flame

    sensor. In this situation it is not possible to continue with flame shut-down

    operations.

    After a minute or two the QTA has cooled sufficiently and the instrument detects

    that the flame has been extinguished. This is not a hazardous situation you must

    merely wait for a minute or two for the QTA to cool before you continue with the

    shut-down procedure.

    Spectrometers with automatic atomizer changeover

    Caution: Risk of damage to the instrument

    If your spectrometer has automatic changeover between the burner system and

    the graphite furnace (e.g. AAnalyst 700 or 800), you can damage the instrument

    if you change to the graphite furnace while the QTA burner mount is installed on

    the burner.

    Before changing atomizers, remove the burner head complete with QTAburner mount from the burner system.