ATOMIC SPECTROMETRY
Directions
Universal QTA Burner Mount
Part Number B300-0350
For Atomic Absorption Spectrometers
a rb dc
Return to Document Menu
-
7/28/2019 Soporte Universal Fias
2/11
Notice
The information contained in this document is subject to change without notice.
Release history
Any comments about the documetation for this product should be addressed to:
User Assistance
PerkinElmer, Inc.
710 Bridgeport Avenue
Shelton, Connecticut 06484-4794
U.S.A.
Or emailed to [email protected]
Notices
The information contained in this document is subject to change without notice. Except as specifically set forth in its
terms and conditions of sale, PerkinElmer makes no warranty of any kind with regaed to this document, including,
but not limited to, the implied warranties of merchantability and fitness for a particular purpose. PerkinElmer shall
not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance
or use of this material.
Copyright information
This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be reproduced in any form whatsoever
or translated into any language without the prior written permission of PerkinElmer, Inc.
or one of its affiliates.
Copyright 1999 by affiliates of PerkinElmer, Inc.
Produced in Singapore
Trademarks
PerkinElmer is a trademark of PerkinElmer, Inc.
AAnalyst trademark of affiliates of PerkinElmer, Inc.
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are not to be consid-
ered unprotected by law.
Part Number Release Publication Date
0993-5222 A
B
C
February 1995
September 1996
November 1999 Valid for Software Version 6.5
-
7/28/2019 Soporte Universal Fias
3/11
0993-5222 3
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5Installing the QTA burner mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Remove the burner head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Attach the QTA burner mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Align the QTA burner mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Reinstall the burner head . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Installing the heat shield . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
QTAs for FIAS systems . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9
Extinguishing the flame . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Spectrometers with automatic atomizer changeover . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
Universal QTA Burner Mount
Marker for header
Contents page
-
7/28/2019 Soporte Universal Fias
4/114 0993-5222
-
7/28/2019 Soporte Universal Fias
5/11
0993-52225
Description
The PerkinElmer Universal QTA Burner Mount is designed to mount the flame-
heated QTA (QTA=quartz tube atomizer) for MHS-10, MHS-15, and FIAS
systems on PerkinElmer atomic absorption spectrometers.
The burner mount installs on the burner head of the AA spectrometer. The QTA
rests in a hinged bracket. When the bracket is in the down position the QTA is
located in the spectrometers sample beam and is heated by the flame. When the
bracket is in the up position the QTA is removed from the sample beam and is not
heated.
Note that the QTA is referred to as the quartz cell or the FIAS-cell in some
publications.
Figure 1. Universal QTA burner mount (shown with MHS-10/15 QTA)
Sample inlet
Hinged QTA bracket(shown in down position)
Burner head
Burner spraychamber
QTA
-
7/28/2019 Soporte Universal Fias
6/11
60993-5222
Installing the QTA burner mount
To install the QTA burner mount you must remove the burner head, attach the
burner mount to the burner head, and then reinstall the burner head.
Remove the burner head
1. Where applicable, extinguish the flame, allow the burner head to cool, shut
down the gases at source, and bleed the gas supply lines (the correct procedure
is described in the Users Guide for your AA spectrometer).
2. Remove the burner head as described in the Users Guide for your AA
spectrometer.
Attach the QTA burner mount
Note: A heat shield is provided with the QTA burner mount. You only requirethe heat shield if your spectrometer is a Model 3100, 3110, or 3300 (see page 8).
1. Swing the QTA bracket on the burner mount into the up position.
2. Slacken (but do not remove) the two adjusting screws in each securing clamp.
3. Slide the QTA burner mount onto the burner head and position the two
securing clamps over the ends of the burner head. The QTA bracket must be
at the front of the burner head.4. Tighten the adjusting screws, but leave them slack enough so that you can
move the burner mount.
Align the QTA burner mount
1. Swing the QTA bracket into the down position.
2. Visually sight along the burner slot and move the QTA burner mount with
respect to the burner head until the two Vs in the QTA bracket line up withthe burner slot.
3. Tighten the adjusting screws in the securing clamps.
W4.2
Warning: High Temperatures Risk of Burns
The burner head can reach very high temperatures.
Do not touch the burner head until it has cooled to room temperature.
-
7/28/2019 Soporte Universal Fias
7/11
0993-5222 7
Figure 2. Attaching the QTA burner mount to the burner head
Figure 3. Aligning the QTA burner mount on the burner head
Reinstall the burner head
1. Reinstall the burner head as described in the Users Guide for your AA
spectrometer.
Note: The QTA burner mount can remain permanently installed on the burner
head.
If you wish to change over to normal flame analyses, tilt the QTA bracket to the
up position and park the QTA in the holder at the side of the MHS-10/15 reactor
assembly. Move the reactor assembly away from the immediate vicinity of the
atomizer compartment.
QTA bracketin up position
Securing clamp(1 of 2)
Adjusting screw(2 each clamp)
Burner head
Burner head
seen from the
rear
Burner slot
V in QTAbracket
QTA bracketin down position
-
7/28/2019 Soporte Universal Fias
8/11
8 0993-5222
Installing the heat shield
If your spectrometer is a Model 3100, 3110, or 3300, you should attach the heat
shield to the burner mount after you have installed the burner mount and
reinstalled the burner head. Refer to Figure 4 while performing the procedure.
Figure 4. Installing the heat shield (Models 3100, 3110, 3300 only)
1. Put the QTA bracket in either the down or the up position.
2. Slide the heat shield onto the QTA bracket so that the slots on the sides of the
heat shield fit on the shoulder screws at each side of the QTA bracket.
Shoulder screw
Heat shield
-
7/28/2019 Soporte Universal Fias
9/11
0993-5222 9
QTAs for FIAS systems
If you are using the flame to heat the QTA for a FIAS system, use only the QTA
with removable PTFE window assemblies. This QTA has a uniform diameter
along its whole length when the window assemblies are removed. Always remove
the window assemblies before heating the QTA.
Do not use a QTA with flared ends and integral exhaust vents.
FIAS QTA suitable forheating with a flame
FIAS QTA unsuitable forheating with a flame
-
7/28/2019 Soporte Universal Fias
10/11
10 0993-5222
Extinguishing the flame
When you extinguish the flame after the system has been in operation for several
hours, the QTA is hot and continues to radiate energy until it has cooled down.
Depending on the model of spectrometer, this energy can be detected by the flame
sensor. In this situation it is not possible to continue with flame shut-down
operations.
After a minute or two the QTA has cooled sufficiently and the instrument detects
that the flame has been extinguished. This is not a hazardous situation you must
merely wait for a minute or two for the QTA to cool before you continue with the
shut-down procedure.
Spectrometers with automatic atomizer changeover
Caution: Risk of damage to the instrument
If your spectrometer has automatic changeover between the burner system and
the graphite furnace (e.g. AAnalyst 700 or 800), you can damage the instrument
if you change to the graphite furnace while the QTA burner mount is installed on
the burner.
Before changing atomizers, remove the burner head complete with QTAburner mount from the burner system.
-